Technical details

Operating mode for plasma ONE Medical device
Rechargeable lithium ion battery: Pack 7,2 V/2,6 Ah
Device output: Maximum 5 W
Device circuit breaker: 1 A polymer built-in/self resetting 30R110 / E183209
Repetition frequency: 420 Hz – 1220 Hz
Converter pulse width: 2,5 μs – 10 μs
Maximum patient current DENTAL: < 100 μA
Patient current MEDICAL: < 100 µA
Automatic power shutdown: t < 0,1 s. max 200 µA
Maximum operating altitude: < 2000 m
Protection class according to EN 60529: IP 20
Operating mode when recharging with power supply (charger)
Power supply input: 100 – 240 V AC (0,8 – 0,4 A)
Isolation voltage: complies with standard EN60950/EN60601
DIN EN 60601-1: 300 V / 1.500 V
Power supply output: 12 V DC / 2,5 A
Dimensions and weight
Length x width x max. height: 22 cm x 19 cm x 8 cm
Weight: 1.250 g
Ambient and storage conditions
Ambient temperature: 15 – 45 °C
Relative humidity: 40 – 85 %
Air pressure: 800 – 1014 hPa
Storage and transport conditions
Ambient temperature: 5 – 60 °C
Relative humidity: 20 – 85 %
Air pressure: 800 – 1014 hPa
Shipping/transport: only in the original case from plasma MEDICAL SYSTEMS® GmbH

Cookies sind Informationen, die von unserem Webserver oder Webservern Dritter an die Web-Browser der Nutzer übertragen und dort für einen späteren Abruf gespeichert werden. Weiterführende Informationen erhalten Sie in der Datenschutzerklärung. Hier finden Sie weitere Informationen zu Cookies. Wenn Sie unseren Service weiterhin nutzen, gehen wir davon aus, dass Sie der Verwendung von Cookies zustimmen.